已注冊用戶需要常規(guī)用戶名密碼登錄并綁定微信,才可以掃碼
點擊右上角可切換登錄模式
還沒有帳號?
現(xiàn)在注冊并享受以下權(quán)益:
- 直達官方商城在線購買
- 請求報價
- 參與精彩問答
- 獲取會員積分
- 兌換禮品
- 和更多 ...
SS7 Verification
SS7,濁度,高量程濁度儀
Perform a calibration
Confirm to perform a baseline using a standardization plate.
The serial number for the last standardization plate used will appear on the display. Confirm to accept or enter the serial number of the plate to be used and confirm.
Follow the controller prompts:
Remove the calibration cylinder and wipe off the top of the sample cylinder.
Place the standardization plate on top of the sample cylinder so that the light beam strikes the center of the plate. Note the orientation of the plate and always place it in the same position when using it to check standardization.
Close the door to eliminate stray light. Confirm to continue.
When the measured value becomes stable, confirm to establish an expected value for the plate.
Note: Future measured values will be compared to the stored expected value. If the established PASS criteria are not met, a new calibration should be performed.
Hach has a complete portfolio of instruments and chemistries with support and services to help you get the right results.
了解更多
隱私政策 Cookie通知 / 滬ICP備13034148號-4 / 滬公網(wǎng)安備31010502004971號 / 滬(浦)應急管危經(jīng)許[2023]201871
? 版權(quán) 2015-2024 哈希中國版權(quán)所有 / 哈希水質(zhì)分析儀器(上海)有限公司 / 本網(wǎng)站用字經(jīng)北大方正電子有限公司授權(quán)許可